The sensitivity of an energy-compensated Geiger-Mueller counter to neutrons with approximate energies of 0.6, 5 and 15 MeV, relative to its sensitivity to gamma rays (ksub(U)-value), has been determined at different centres in Europe. The time-of-flight, angular difference, associated-particle coincidence and lead filtration technique have been compared, using the same Geiger-Mueller counter. For 15 MeV neutrons, a consistent ksub(U)-value of 1.9% was obtained. For the lower neutron energies, lower ksub(U) values were observed with reasonable agreement. (author)