Published June 1, 1977
| Version v1
Journal article
High-energy gamma spectra detected with imporved HgI2 spectrometers at room temperature
- 1. The Hebrew University of Jerusalem, Israel
Description
The detection of the photopeak of the 662-keV γ ray of 137Cs with a full width at half-maximum (FWHM) of only 6.5 keV by a 0.5-mm-thick improved stoichiometry HgI2 crystal is reported. The 1173- and 1332-keV γ photopeaks of 60Co have also been resolved for the first time. Polarization effects were negligible
Additional details
Identifiers
- DOI
- 10.1063/1.89256;
Publishing Information
- Journal Title
- Applied Physics Letters
- Journal Volume
- 30
- Journal Issue
- 11
- Series
- Appl. Phys. Lett.
- Journal Page Range
- 611-613
- ISSN
- 0003-6951
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 8328480
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- AMERICIUM 241; CESIUM 137; COBALT 57; COBALT 60; CRYSTAL GROWTH; ENERGY RESOLUTION; FABRICATION; FLUORESCENCE; GAMMA DETECTION; GAMMA DOSIMETRY; GAMMA SPECTRA; GAMMA SPECTROMETERS; GAMMA SPECTROSCOPY; HEAT TREATMENTS; HGI2 SEMICONDUCTOR DETECTORS; KEV RANGE
- Descriptors DEC
- ACTINIDE NUCLEI; ALPHA DECAY RADIOISOTOPES; AMERICIUM ISOTOPES; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; COBALT ISOTOPES; DAYS LIVING RADIOISOTOPES; DOSIMETRY; ELECTRON CAPTURE RADIOISOTOPES; ENERGY RANGE; HEAVY NUCLEI; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTO; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LUMINESCENCE; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTION; RADIATION DETECTORS; RADIOISOTOPES; RESOLUTION; SEMICONDUCTOR DETECTORS; SPECTRA; SPECTROMETERS; SPECTROSCOPY; YEARS LIVING RADIOISOTOPES