Published December 15, 1986
| Version v1
Journal article
Direct measurement of the energy spectrum of an intense proton beam
- 1. Sandia National Laboratories, Albuquerque, New Mexico 87185
Description
A time-resolved magnetic spectrograph was used to measure the energy spectrum of an intense (0.7 TW/cm2) proton beam. A thin gold foil placed at the focus of an ion-diode Rutherford-scattered protons by 900 into the spectrograph, reducing the beam intensity to a level suitable for magnetic analysis. The scattered protons were collimated and then deflected in a samarium-cobalt permanent magnet. The deflected protons were recorded simultaneously on CR-39 and eight p-i-n diodes. A Monte Carlo computer code was used to calculate the sensitivity and resolution of the spectrograph. Data taken on the Proto I accelerator shows a 150--250-keV-wide proton-energy spectrum at each instant in time
Additional details
Publishing Information
- Journal Title
- J. Appl. Phys.
- Journal Volume
- 60
- Journal Issue
- 12
- Series
- J. Appl. Phys.
- Journal Page Range
- 4059-4063
- ISSN
- 0021-8979
- CODEN
- JAPIA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18023572
- Subject category
- S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- ENERGY SPECTRA; KEV RANGE 100-1000; MAGNETIC SPECTROMETERS; MONTE CARLO METHOD; PROTON BEAMS; RESOLUTION; RUTHERFORD SCATTERING; SENSITIVITY
- Descriptors DEC
- BEAMS; ELASTIC SCATTERING; ENERGY RANGE; KEV RANGE; MEASURING INSTRUMENTS; NUCLEON BEAMS; PARTICLE BEAMS; SCATTERING; SPECTRA; SPECTROMETERS