Evaluation of a Commercial 4-D Diode Array for Helical Tomotherapy Plan Verification
Creators
- 1. Department of Medical Physics, National Institute for Cancer Research, Genoa (Italy)
Description
The performance of a commercial 4-D diode array, the ArcCHECK (Sun Nuclear, FL, USA) has been tested for helical tomotherapy patient-specific quality assurance (QA). Seventeen helical tomotherapy plans were selected representing a variety of tumour sites. Agreement between measured and calculated dose distributions was assessed by evaluating the gamma index (γ) with criteria ranging from 2% dose differences or 2 mm distance to agreement (2%/2 mm) to 3%/3 mm. For a subset of plans, some delivery errors were voluntarily introduced in all the four rotational-translational directions to investigate the identification ability of the device. For a single plan point, dose comparison was carried out in the low dose gradient region in order to quantify the amount of dose deviation as a function of the absolute dose delivered. QA plans were also calculated and delivered to radiochromic films (GafChromic EBT2, ISP Corp., USA) within a cylindrical solid water phantom for comparison. For the 16 plans delivered, the mean percentage value of points with γ < 1 was 97.2% ± 2.2%, 94.2% ± 4.6%, 95.1% ± 3.8% and 90.8% ± 6.2% calculated with (3%/3 mm), (3%/2 mm), (2%/3 mm) and (2%/2 mm) criteria, respectively, demonstrating that the 4-D diode array can be used for rotational IMRT QA. Four of the 'wrong' measurements still had a good QA response showing a number of points >90% fulfilling γ < 1 calculated with the 3%/3 mm criterion. The number of 'false positives' decreased with tighter γ calculations (3%/2 mm and 2%/3 mm) and disappeared with 2%/2 mm. Point to point comparison showed a mean dose deviation of 1.2% ± 1.1%, 1.9% ± 1.6% and 2.1% ± 1.3% for low, medium and high dose regions, respectively. The mean percentage value of points with γ < 1 was 96.3% ± 3.3% and 91.4% ± 2.4% calculated with (3%/3 mm) for diode array and film dosimetry showing a consistent superiority of the device with respect to GafChromic film (p < 0.05). (author)
Additional details
Identifiers
Publishing Information
- Publisher
- IAEA
- Imprint Place
- Vienna (Austria)
- ISBN
- 978-92-0-116210-6
- Imprint Title
- Standards, Applications and Quality Assurance in Medical Radiation Dosimetry (IDOS). Proceedings of an International Symposium. V. 2
- Imprint Pagination
- 495 p.
- Series
- Proceedings Series
- Journal Page Range
- p. 237-243
- ISSN
- 0074-1884
Conference
- Title
- International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry
- Acronym
- IDOS 2010
- Dates
- 9-12 Nov 2010
- Place
- Vienna (Austria)
INIS
- Country of Publication
- Austria
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43009162
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; DOSES; FILM DOSIMETRY; FILMS; QUALITY ASSURANCE; RADIATION DOSE DISTRIBUTIONS; USA
- Descriptors DEC
- DEVELOPED COUNTRIES; DIAGNOSTIC TECHNIQUES; DOSIMETRY; MEDICINE; NORTH AMERICA; NUCLEAR MEDICINE; RADIOLOGY; RADIOTHERAPY; THERAPY; TOMOGRAPHY
Optional Information
- Notes
- 1 fig., 2 tabs., 8 refs. Imprint:In two volumes; A CD-ROM of the contributed papers accompanies these proceedings.
- Secondary number(s)
- STI/PUB--1514(vol.2); IAEA-CN--182