Patient doses from neuroradiological procedures
Creators
- 1. Department of Human Services, Melbourne, VIC (Australia). Radiation Safety Unit
- 2. The Royal Melbourne Hospital, VIC (Australia). Department of Radiology
Description
Patient doses have been determined for a sample of 20 patients in a hospital department undergoing various interventional neuroradiological procedures. Skin entrance dose, dose-area product and effective dose have been estimated. Effective dose was estimated from published conversion factors. The highest dose on the patients body was recorded as the maximum skin entrance dose. The maximum skin entrance dose varied from 0.2 to 2.3 Gy. The largest maximum skin entrance dose is similar to the threshold for skin injuries (erythema). The effective dose varied from 2 to 34 mSv. The results of this study are compared with the results of other studies on radiation dose to patients from neuroradiological procedures. Copyright (1999) Australasian Physical and Engineering Sciences in Medicine and the College of Biomedical Engineers
Additional details
Publishing Information
- Journal Title
- Australasian Physical and Engineering Sciences in Medicine
- Journal Volume
- 22
- Journal Issue
- 4
- Journal Page Range
- p. 153-162
- ISSN
- 0158-9938
- CODEN
- AUPMDI
INIS
- Country of Publication
- Australia
- Country of Input or Organization
- Australia
- INIS RN
- 31037835
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Numerical Data
- Descriptors DEI
- BIOLOGICAL RADIATION EFFECTS; BIOMEDICAL RADIOGRAPHY; DOSIMETRY; EXPERIMENTAL DATA; HEAD; LOCAL RADIATION EFFECTS; NECK; PATIENTS; POST-IRRADIATION EXAMINATION; RADIATION DOSES; RADIATION INJURIES; SKIN
- Descriptors DEC
- ANIMALS; BIOLOGICAL EFFECTS; BIOLOGICAL RADIATION EFFECTS; BODY; DATA; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; INFORMATION; INJURIES; MAMMALS; MAN; MEDICINE; NUCLEAR MEDICINE; NUMERICAL DATA; ORGANS; PRIMATES; RADIATION EFFECTS; RADIOLOGY; VERTEBRATES
Optional Information
- Notes
- 22 refs., 7 tabs.,