Shielding Design and Leaking Measurement for the High Energy Radiation Room
Description
An optimum shielding design and the computation of protective barriers for high energy radiation therapy room, Toshiba 13 MeV , are presented. We obtained following results by comparison between the p recalculating values and actual survey after complete installation of radio generating units. 1. The precalculating values of protective barrier are 5 times more protective than that of actual measurement. 2. The dose rate during exposure are 2-10 mR /hr at out of the door and the control room. 3. The exposure doses for occupationally persons are relatively low levels, the average values of exposure dose is 10-50 mR per month. 4. The foul smelling and ozone gas production from long exposure of cancer patients cannot be eliminated when the room is ill ventilated