3T MRI of the knee with optimised isotropic 3D sequences. Accurate delineation of intra-articular pathology without prolonged acquisition times
Creators
- 1. University College Dublin, Radiography and Diagnostic Imaging, School of Medicine, Belfield, Dublin (Ireland)
- 2. University College Dublin, School of Medicine, Dublin (Ireland)
- 3. Mater Misericordiae University Hospital, Department of Radiology, Dublin (Ireland)
- 4. Mater Misericordiae University Hospital, Department of Orthopaedics, Dublin (Ireland)
Description
To investigate optimised isotropic 3D turbo spin echo (TSE) and gradient echo (GRE)-based pulse sequences for visualisation of articular cartilage lesions within the knee joint. Optimisation of experimental imaging sequences was completed using healthy volunteers (n=16) with a 3-Tesla (3T) MRI scanner. Imaging of patients with knee cartilage abnormalities (n=57) was then performed. Acquired sequences included 3D proton density-weighted (PDW) TSE (SPACE) with and without fat-suppression (FS), and T2*W GRE (TrueFISP) sequences, with acquisition times of 6:51, 6:32 and 5:35 min, respectively. One hundred sixty-one confirmed cartilage lesions were detected and categorised (Grade II n=90, Grade III n=71). The highest sensitivity and specificity for detecting cartilage lesions were obtained with TrueFISP with values of 84.7% and 92%, respectively. Cartilage SNR mean for PDW SPACE-FS was the highest at 72.2. TrueFISP attained the highest CNR means for joint fluid/cartilage (101.5) and joint fluid/ligament (156.5), and the lowest CNR for cartilage/meniscus (48.5). Significant differences were identified across the three sequences for all anatomical structures with respect to SNR and CNR findings (p-value <0.05). Isotropic TrueFISP at 3T, optimised for acquisition time, accurately detects cartilage defects, although it demonstrated the lowest contrast between cartilage and meniscus. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-017-4816-xAdditional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 27
- Journal Issue
- 11
- Journal Page Range
- p. 4563-4570
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 49004563
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BONE JOINTS; CARTILAGE; CORRECTIONS; FAT CELLS; FEMUR; IMAGE PROCESSING; MAGNETIC FIELDS; MALFORMATIONS; NMR IMAGING; NOISE; RELAXATION TIME; S CODES; THREE-DIMENSIONAL CALCULATIONS; TIBIA; WEIGHTING FUNCTIONS
- Descriptors DEC
- ANIMAL CELLS; ANIMAL TISSUES; BODY; COMPUTER CODES; CONNECTIVE TISSUE; CONNECTIVE TISSUE CELLS; DIAGNOSTIC TECHNIQUES; FUNCTIONS; ORGANS; PATHOLOGICAL CHANGES; PROCESSING; SKELETON; SOMATIC CELLS