Implementation and quantitative evaluation of analytical methods for attenuation correction in SPECT: a phantom study
Creators
- 1. Department of Radiology, Showa University Fujigaoka Hospital, 1-30 Fujigaoka Aoba-ku, Yokohama 227-0043 (Japan)
- 2. Department of Information Processing, Yokohama Soei College, 1 Mihocho Midori-ku, Yokohama 226-0016 (Japan)
- 3. Department of Research for Nuclear Medicine, Medical Systems Division, Shimadzu Corporation, 1 Nishinokyo-Kuwabarocho Nakagyo-ku, Kyoto 604-8442 (Japan)
Description
Three differing exact methods of inverting the two-dimensional (2D) exponential Radon transform were implemented and evaluated quantitatively with a phantom study. The phantom had the shape of a pie-chart divided into six cavities, each 480 ml in volume and 10 cm in height, that were symmetrically positioned in a cylinder that was 20 cm in diameter and 10 cm in height. This phantom tests for linearity between true activity concentration and measured activity concentration, and it is denoted as a linearity phantom in the present study. Each cavity contained a different concentration of a homogeneous solution of 99mTc (74, 148, 222, 296, 370 and 444 kBq ml-1). Data acquisition was performed with two energy windows: a 20% photopeak energy window set symmetrically over the 140 keV of 99mTc and a secondary 5% energy window set over the 122 keV peak. We optimized a triple-energy window scatter correction method for a gamma camera-collimator system to obtain accurate scatter-corrected projections. A circular ROI 3 cm in diameter was identified over each cavity region, and count density (counts per pixel) was calculated. This value was converted to activity concentration (kBq ml-1) using a cross-calibration coefficient between SPECT counts and the gamma well counter. The relation between true activity (x) and measured activity concentration (y) was fitted to a line using the least-squares method. Regression lines were y=0.63+1.0255x (R2=0.9987), y=-2.62+1.0278x (R2=0.9995), and y=0.092+1.0241x (R2=0.9989) for the Bellini, Inouye and Metz-Pan methods respectively. In another phantom study using two different types of phantoms, contrast of a cold region in the two was 96% and 101% for all three methods. Combined optimized scatter correction and analytical attenuation correction methods achieve good accuracy in quantification of activity distribution with a uniform attenuating medium. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology (Online)
- Journal Volume
- 44
- Journal Issue
- 10
- Journal Page Range
- p. 2643-2655
- ISSN
- 1361-6560
INIS
- Country of Publication
- International Atomic Energy Agency (IAEA)
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 31025391
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; ATTENUATION; CALCULATION METHODS; CORRECTIONS; EVALUATION; PHANTOMS; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; MOCKUP; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- Refs