Published March 1977 | Version v1
Journal article

Sumpeak calibration of 123I

  • 1. Instituut voor Kernphysisch Onderzoek, Amsterdam (Netherlands)

Description

The sumpeak method applied to the calibration of 123I is discussed. The decay scheme of 123I and the spectra for 123I, measured with a 60 x 60 mm NaI(Tl) crystal, as used in the calculations are shown. Necessary corrections are considered. The conclusion is that the sumpeak method for 123I is only useful in the case of high geometrical efficiency, i.e. inside the well of a NaI(Tl) detector. Thus, as is demonstrated, the method is even applicable to solutions and this is considered a clear advantage. (U.K.)

Additional details

Identifiers

Publishing Information

Journal Title
The International Journal of Applied Radiation and Isotopes
Journal Volume
28
Journal Issue
3
Series
Int. J. Appl. Radiat. Isot.
Journal Page Range
271-275
ISSN
0020-708X

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent