A clinical evaluation of visual feedback-guided breath-hold reproducibility of tumor location
Creators
- 1. Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, Maidashi 3-1-1, Higashi-ku, Fukuoka 812-8582 (Japan)
- 2. Department of Radiology, Kyushu University Hospital, Beppu, 4546 Tsurumibara, Beppu-shi, Oita 874-0838 (Japan)
- 3. Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, Maidashi 3-1-1, Higashi-ku, Fukuoka 812-8582 (Japan)
Description
The purpose of this study was to evaluate the reproducibility of visual feedback-guided breath-hold using a machine vision system with a charge-coupled device camera and a monocular head-mounted display. Sixteen patients with lung tumors who were treated with stereotactic radiotherapy were enrolled. A machine vision system with a charge-coupled device camera was used for monitoring respiration. A monocular head-mounted display was used to provide the patient with visual feedback about the breathing trace. The patients could control their breathing so that the breathing waveform would fall between the upper and lower threshold lines. Planning and treatment were performed under visual feedback-guided expiratory breath-hold. Electronic portal images were obtained during treatment. The range of cranial-caudal motion of the tumor location during each single breath-hold was calculated as the intra-breath-hold (intra-BH) variability. The maximum displacement between the two to five averaged tumor locations of each single breath-hold was calculated as the inter-breath-hold (inter-BH) variability. All 16 patients tolerated the visual feedback-guided breath-hold maneuvers well. The intra- and inter-BH variability of all patients was 1.5 ± 0.6 mm and 1.2 ± 0.5 mm, respectively. A visual feedback-guided breath-hold technique using the machine vision system is feasible with good breath-hold reproducibility.
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/54/23/009Additional details
Identifiers
- DOI
- 10.1088/0031-9155/54/23/009;
- PII
- S0031-9155(09)27759-1;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 54
- Journal Issue
- 23
- Journal Page Range
- p. 7171-7182
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41062297
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BREATH; CAMERAS; CHARGE-COUPLED DEVICES; FEEDBACK; IMAGES; LUNGS; MONITORING; NEOPLASMS; RADIOTHERAPY; RESPIRATION; VISION
- Descriptors DEC
- BODY; DISEASES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RESPIRATORY SYSTEM; SEMICONDUCTOR DEVICES; THERAPY