Published 1987
| Version v1
Journal article
Feasibility study of using CaSO4:Dy phosphor for simultaneous estimation of exposure and time elapsed post-exposure
Creators
- 1. National Tsing Hua Univ., Hsinch, Taiwan, Republic of China
Description
Samples of locally made CaSO4:Dy phosphor were investigated to study the feasibility of simultaneous estimation of exposure and time elapsed post-exposure. The glow curve behaviour relative to absorbed dose, radiation energy, ambient temperature and time were studied in detail. An attempt was made using the variation of the glow peak ratio P2/P1 to estimate the time elapsed after an accident exposure. Results indicate that the limitation for the successful estimation of this elapsed time seems to be the temperature sensitivity of the peak ratios. (author)
Additional details
Publishing Information
- Journal Title
- Radiat. Prot. Dosim.
- Journal Volume
- 18
- Journal Issue
- 3
- Series
- Radiat. Prot. Dosim.
- Journal Page Range
- 157-161
- ISSN
- 0144-8420
- CODEN
- RPDOD
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 18095704
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALCIUM SULFATES; FEASIBILITY STUDIES; GLOW CURVE; PHOSPHORS; THERMOLUMINESCENT DOSEMETERS; TIME DEPENDENCE
- Descriptors DEC
- ALKALINE EARTH METAL COMPOUNDS; CALCIUM COMPOUNDS; DOSEMETERS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; OXYGEN COMPOUNDS; SULFATES; SULFUR COMPOUNDS