PREDICTION OF THE CORROSION RESISTANCE OF Ti-Ni-Cr ALLOY IN PHYSIOLOGICAL SOLUTION

  • 1. F. Tavadze Institute of Metallurgy and Materials Science, Tbilisi (Georgia)

Description

The prediction of long-term corrosive behavior of Ti-Ni-Cr alloy was carried out at 37 ℃ in the physiological solution (0.9% NaCl) with application of electrochemical and corrosion tests, spectral chemical analysis and statistical methods of experimental data analysis. The obtained equation of time dependence of the corrosion rate is expressed as C = 0.0038 exp (-0.045τ). The time dependences of corrosion potentiasls and rates and the absence of pitting corrosion allows forecasting the high corrosion resistance of alloy corresponding to the passive state during long-term exposure. The chemical spectral analysis confirmed the calculated and predicted low values of corrosion rates and showed that, during the long-term exposure, the quantity of titanium ions passed into the solution could be neglected. (author)

Additional details

Publishing Information

Journal Title
Georgian Engineering News (GEN)
Journal Issue
n.4,2008
Journal Page Range
p. 53-57
ISSN
1512-0287

Optional Information

Notes
7 refs., 2 figs., 2 tabs.