Published 2009 | Version v1
Book

Multichannel dosimeter and α-Al2O3:C Optically Stimulated Luminescence (OSL) fiber sensors for use in radiation therapy - Evaluation with photon beams

  • 1. CEA, LIST, Laboratoire de Mesures Optiques, F-91191 Gif-sur-Yvette (France)
  • 2. CEA, LIST, Laboratoire National Henri Becquerel, F-91191 Gif-sur-Yvette (France)
  • 3. Institut Gustave Roussy, Service de Physique, 39 rue Camille Desmoulins, F-94805 Villejuif (France)

Description

A multichannel OSL Fibre Optic dosimeter based on Optically Stimulated Luminescence (OSL) of alumina is proposed for on-line In Vivo Dosimetry (IVD) in Radiation Therapy (RT). Two types of dosimetric-grade Al2O3:C crystals are compared and show different behavior according to manufacturing process. Metrological validations have been performed with a Saturne 43 LINAC in reference conditions at CEA LIST LNHB (French Ionizing Radiation Reference Laboratory). The dose response of OSL integrals under photon beam irradiation (6, 12 and 20 MV) show sub-linearity behavior modeled by second-order equations and exhibit a small energy dependence (between 0.7% and 1.4%), explained by a modified intermediate cavity model adapted to a Linac photon spectrum. Preclinical tests at Institut Gustave Roussy (IGR) prove that a proper design for a PMMA build-up cap leads to a low dependence vs photon beam orientation (± 1.5% and ± 0.9 %) and vs field size in view of surface measurements. (authors)

Availability note (English)

Available from doi:

Additional details

Identifiers

Publishing Information

Publisher
IEEE - Institute of Electrical and Electronics Engineers
Imprint Place
New York (United States)
ISBN
978-1-4244-5207-1
Imprint Pagination
10 p.

Conference

Title
1. International Conference on Advancements in Nuclear Instrumentation, Measurement Methods and their Applications
Acronym
ANIMMA 2009
Dates
7-10 Jun 2009
Place
Marseille (France)

Optional Information

Notes
25 refs.