Published November 1986 | Version v1
Journal article

Determination of selenium and tellurium in the gas phase using specific columns and atomic absorption spectrometry

  • 1. Univ. of Missouri, Rolla

Description

Total selenium and tellurium in the gas phase were analyzed after adsorption on gold-coated beads and charcoal. The thermally eluting gas was trapped on columns filled with quartz beads that were cooled in an ice bath. The beads were boiled in dilute HCl, and the resulting solution was analyzed for Se and Te by graphite furnace atomic absorption spectrometry. Their results demonstrate that gold-coated beads efficiently trap gaseous Se and Te at a low gas flow rate, but at higher flow rates charcoal traps are more expedient. With charcoal traps, it was found that local air samples contain Se in the range of 0.92-3.05 ng m-3 and Te in the range of 0.10-0.34 ng m-3. Detection limits down to about 0.1 ng m-3 allow the ready detection of Se and Te in rural air with a precision of about +/- 6% at the nanogram level of Te and about +/- 4% at the nanogram level of Se

Additional details

Publishing Information

Journal Title
Anal. Chem. (Wash.)
Journal Volume
58
Journal Issue
13
Series
Anal. Chem. (Wash.).
Journal Page Range
2811-2813
ISSN
0003-2700
CODEN