Published August 1979 | Version v1
Journal article

A source storage and irradiator device for calibrating direct-reading pocket dosimeters

  • 1. Aerospace Medical Div., Brooks AFB, TX (USA)

Description

To reduce personnel radiation exposure during the calibration of direct-reading pocket dosimeters, a cable operated irradiator device was developed for handling and storing a 9.76 mg radium needle. This device is located inside a hot cell which includes remotely operated manipulators. It offers excellent shielding, simple construction and operation with no exposure to personnel during calibration. (author)

Additional details

Publishing Information

Journal Title
Health Phys.
Journal Volume
37
Journal Issue
2
Series
Health Phys.
Journal Page Range
244-247
ISSN
0017-9078