Experimental study of neutron dose measurements with CR-39 detector
Creators
- 1. Liaoning Inst. of Radiation Hygiene and Protect, Shenyang (China)
- 2. Chinese Academy of Measure Sciences, Beijing (China)
- 3. Liaoning Inst. of Labor Hygiene and Occupational Diseases Prevention, Shenyang (China)
Description
The author describes the measurements of neutron dose using CR-39 detectors, its method, etching condition and results of field application. The optimum etching condition found by orthogonal experimental design was 6.5 mol/L KOH solution, at 70 degree C 7 hours. The calibration factor was 7.0 x 10-3 mSv/tr·cm-2. The detectable limit was 0.06 mSv