Description

The relative detection efficiency of neutron long counter was calibrated with 241Am-Be neutron source. Energy conversion coefficient between 5.1 MeV and 14 MeV was also determined by the associated particle method. At the same time, the relative detection efficiency and energy conversion coefficient were calculated with Monte-Carlo method. In present work, the uncertainty of relative detection efficiency from experiment and calculation is 10.8%0 and 2.3%, respectively. Both are coincident in the range of uncertainty. (authors)

Additional details

Publishing Information

Journal Title
Atomic Energy Science and Technology
Journal Volume
42
Journal Issue
2
Journal Page Range
p. 177-179
ISSN
1000-6931

Optional Information

Notes
2 tabs., 5 refs.