SU-E-T-223: Investigation of the Accuracy of Two-Dimensional Dose Distributions Measurement From High-Dose-Rate Brachytherapy Ir-192 Source Using Multiple-Diode-Array Detector (MapCheck2)
- 1. Oklahoma Univ. Health Science Ctr., Oklahoma City, OK (United States)
Description
Purpose: To investigate the dosimetric accuracy of multiple-diode-array detector (Mapcheck2) for high-dose-rate brachytherapy Ir-192 source. The two-dimensional (2D) dose distributions measured with MapCheck2 were validated with EBT2 Gafchromic film measurement and AAPM task-group- 43 (TG-43) modeling. Methods: 2D-dose distributions from Ir-192 source were measured with MapCheck2 and EBT2-films. MapCheck2 response was corrected for effects: directional dependence, diode and phantom heterogeneity. Optical density growth of the film was controlled by synchronized scanning of the film exposed to Ir-192 and calibration films exposed to 6 MV linac beams. Similarly, MapCheck2 response was calibrated to dose using 6 MV beams. An empirical model was developed for the dose distributions measured with Mapcheck2 that considered directional, diode and phantom heterogeneity corrections. The dose deposited in solid-state-detectors was modeled using a cavity theory model for the diode. This model was then validated with measurements using EBT2-films and calculations with TG-43. Results: The response of MapCheck2 has been corrected for different effects including: (a) directional dependence of 0–20% over angular range 0o–90o, (b) phantom heterogeneity (3%) and (c) diode heterogeneity (9%). The corrected dose distributions measured with MapCheck2 agreed well with the measured dose distributions from EBT2-film and with calculations using TG-43 within 5% over a wide range of dose levels and rates. The advantages of MapCheck2 include less noisy, linear and stable response compared with film. The response of MapCheck2 exposed to 192Ir-source showed no energy dependence similar to its response to MV energy beam. Detection spatial-resolution of individual diodes was 0.8×0.8 mm2, however, 2DMapCheck2 resolution is limited by distance between diodes (7.07 mm). Conclusion: The dose distribution measured with MapCheck2 agreed well within 5% with that measured using EBT2-films; and calculations with TG- 43. Considering correction of artifacts, MapCheck2 provides a compact, practical and accurate dosimetric tool for measurement of 2D-dose distributions for brachytherapy Ir-192
Additional details
Identifiers
- DOI
- 10.1118/1.4888553;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 41
- Journal Issue
- 6
- Journal Page Range
- p. 274
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46057397
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- ACCURACY; BRACHYTHERAPY; CORRECTIONS; DOSE RATES; FILM DOSIMETRY; IRIDIUM 192; LINEAR ACCELERATORS; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; SPATIAL RESOLUTION
- Descriptors DEC
- ACCELERATORS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; DAYS LIVING RADIOISOTOPES; DOSES; DOSIMETRY; ELECTRON CAPTURE RADIOISOTOPES; HEAVY NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IRIDIUM ISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEDICINE; MINUTES LIVING RADIOISOTOPES; MOCKUP; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; RADIOISOTOPES; RADIOLOGY; RADIOTHERAPY; RESOLUTION; STRUCTURAL MODELS; THERAPY; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- (c) 2014 American Association of Physicists in Medicine