Over-response of tissue-equivalent ionization chambers at low energies: effect of high-Z surface contamination

  • 1. Department of Radiology, Aultman Hospital, Canton, Ohio

Description

Inner wall surface contamination of an ionization chamber by high-Z materials can produce an over-response to low-energy photons which is greatly underestimated by the percentage composition of the contaminant in the entire chamber. Experience with a chamber contaminated in such a manner is presented. Examples are included which illustrate the magnitude of possible clinical effects from low-energy units which have been calibrated with a chamber contaminated with a thin layer of relatively high-Z material

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
11
Journal Issue
1
Series
Med. Phys.
Journal Page Range
65-66
ISSN
0094-2405

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
15064357
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
ERRORS; IONIZATION CHAMBERS; PHOTONS; SURFACE CONTAMINATION; TISSUE-EQUIVALENT MATERIALS
Descriptors DEC
CONTAMINATION; ELEMENTARY PARTICLES; MASSLESS PARTICLES; MATERIALS; MEASURING INSTRUMENTS; RADIATION DETECTORS