Published 1976 | Version v1
Book

Precision, accuracy, and fading characteristics of calcium fluoride dysprosium activated thermoluminescent dosimeters (CaF2: Dy TLD'S)

Description

CaF2:Dy TLD-200, G-2 cards were evaluated to determine their suitability for measuring background radiation. All laboratory exposures utilized an automatic strontium -90 Beta source calibration unit. This calibration unit is part of the Harshaw Model 2271 Automated Personnel Monitoring System which was used in the analysis of the G-2 cards. The parameters evaluated were: (1) precision and accuracy; (2) reproducibility; (3) linearity of response; and (4) fading

Additional details

Publishing Information

Publisher
Rensselaer Polytechnic Institute.
Imprint Place
Troy, NY
Imprint Title
Tenth midyear topical symposium of the Northeastern New York Chapter, Health Physics Society, on natural radioactivity in man's environment
Journal Page Range
p. 546-554.

Conference

Title
10. midyear symposium for Health Physics Society.
Dates
11 - 13 Oct 1976.
Place