Average glandular dose in routine mammography screening using a Sectra Microdose Mammography unit
- 1. Dept. of Medical Radiation Physics, Malmoe Univ. Hospital, SE-205 02 Malmoe (Sweden)
- 2. RTI Electronics AB, SE-431 37 Moelndal (Sweden)
- 3. Dept. of Diagnostic Radiology, Malmoe Univ. Hospital, SE-205 02 Malmoe (Sweden)
- 4. Dept. of Diagnostic Radiology, Lund Univ. Hospital, SE-221 85 Lund (Sweden)
- 5. Dept. of Diagnostic Radiology, Hospital of Helsingborg, SE-251 87 Helsingborg (Sweden)
Description
The Sectra MicroDose Mammography system is based on direct photon counting (with a solid-state detector), and a substantially lower dose to the breast than when using conventional system can be expected. In this work absorbed dose measurements have been performed for the first unit used in routine mammography screening (at the Hospitals of Helsingborg (Sweden)). Two European protocols on dosimetry in mammography have been followed. Measurement of half value layer (HVL) cannot be performed as prescribed, but this study has demonstrated than non-invasive measurements of HVL can be performed accurately with a sensitive and well collimated solid-state detector with simultaneous correction for the energy dependence. The average glandular dose for a 50 mm standard breast with 50% glandularity, simulated by 45 mm polymethylmethacrylate, was found to be 0.21 and 0.28 mGy in March and December 2004, respectively. These values are much lower than for any other mammography system on the market today. It has to be stressed that the measurement were made using the current clinical settings and that no systematic optimisation of the relationship between absorbed dose and diagnostic image quality has been performed within the present study. In order to further increase the accuracy of absorbed dose measurements for this unit, the existing dose protocols should be revised to account also for the tungsten/aluminium anode/filter combination, the multi-slit pre-collimator device and the occurrence of a dose profile in the scanning direction. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/nch556Additional details
Identifiers
- DOI
- 10.1093/rpd/nch556;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 114
- Journal Issue
- 1-3
- Journal Page Range
- p. 436-443
- ISSN
- 0144-8420
Conference
- Title
- 2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
- Dates
- 23-25 Apr 2004
- Place
- Malmoe (Sweden)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 44035837
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; ALUMINIUM; BIOMEDICAL RADIOGRAPHY; DOSIMETRY; ENERGY DEPENDENCE; HOSPITALS; IMAGES; MAMMARY GLANDS; OPTIMIZATION; PHOTONS; RADIATION DETECTORS; RADIATION DOSES; SCREENING; SIMULATION; STANDARDS; TUNGSTEN; X RADIATION
- Descriptors DEC
- BODY; BOSONS; BUILDINGS; DIAGNOSTIC TECHNIQUES; DOSES; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; ELEMENTS; GLANDS; IONIZING RADIATIONS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MEDICAL ESTABLISHMENTS; MEDICINE; METALS; NUCLEAR MEDICINE; ORGANS; RADIATIONS; RADIOLOGY; REFRACTORY METALS; TRANSITION ELEMENTS
Optional Information
- Notes
- 19 refs