Published May 1989
| Version v1
Journal article
On the use of the residual range as the only energy parameter in electron dosimetry
Creators
- 1. Physikalisch-Technische Bundesanstalt, Braunschweig (Germany, F.R.)
Description
A commonly used procedure to determine the absorbed dose to water in a high energy electron beam is based on the assumption that the response of an air-filled plane parallel chamber for this quantity is only a function of the difference between practical range and measuring depth (the residual range). This procedure is said to be applicable to almost all electron energies and spectra in practical use. The response of a plane parallel chamber in a water phantom was determined by means of ferrous sulphate dosimeters in radiation fields of a linear accelerator. The results show that the residual range is insufficient as the only energy parameter. (orig.)
Abstract (German)
Ein gebraeuchliches Verfahren zur Bestimmung der Wasser-Energiedosis in einem Wasserphantom fuer hochenergetische Elektronenstrahlung ist an die Voraussetzung gebunden, dass das Ansprechvermoegen einer luftgefuellten Flachkammer fuer diese Messgroesse allein eine Funktion der Differenz aus praktischer Reichweite und Messtiefe (der sogenannten Restreichweite) ist. Das Verfahren soll in einem grossen Energiebereich fuer alle in der Praxis vorkommenden Elektronenspektren anwendbar sein. Durch eine experimentelle Bestimmung des Ansprechvermoegens einer Flachkammer im Wasserphantom mit Hilfe von Eisensulfatdosimetern in Strahlungsfeldern eines Linearbeschleunigers wird gezeigt, dass die Restreichweite als alleiniger Energieparameter nicht ausreichend ist. (orig.)Additional details
Additional titles
- Original title (German)
- Zur Verwendung der Restreichweite als alleinigen Energieparameter in der Elektronendosimetrie
Publishing Information
- Journal Title
- Strahlentherapie und Onkologie
- Journal Volume
- 165
- Journal Issue
- 5
- Series
- Strahlenther. Onkol.
- Journal Page Range
- 392-394
- ISSN
- 0179-7158
- CODEN
- STONE
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 20060519
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ELECTRON DOSIMETRY; ENERGY DEPENDENCE; IONIZATION CHAMBERS; PHANTOMS; RADIOTHERAPY; SENSITIVITY; WATER
- Descriptors DEC
- DOSIMETRY; HYDROGEN COMPOUNDS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; OXYGEN COMPOUNDS; POLAR SOLVENTS; RADIATION DETECTORS; SOLVENTS; STRUCTURAL MODELS; THERAPY