Coronary cineangiography and ionizing radiation exposure to patients: analysis of primary and secondary beam

  • 1. Universidad de Chile, Santiago (Chile). Hospital Clinico
  • 2. Comision Chilena de Energia Nuclear (CCHEN), Santiago (Chile)

Description

The purpose of this work was to determine the level of exposure dose to patients during coronariographies in different areas of body. This study has presented the medical surveillance of 18 cases and the radiation monitoring of these patients by TLD in thyroid and pelvis (secondary beam) and, in the right and left scapular region (primary beam) for each one of these procedures. The ionizing radiation received was 215 ± 200 mGy in left scapular region (range 1-710) and 255±213 mGy in the right scapular region (range 22-635) p=NS. In the pelvic region the ionizing radiation was 0,22±0,06 mGy and in the thyroid region was 3,62±2,44 mGy

Part of:
Proceedings of the 5. Regional congress on radiation protection and safety; 2. Iberian and Latin American Congress on Radiological Protection Societies; Regional IRPA Congress

Additional details

Additional titles

Original title (Spanish)
Cineangiografia coronaria y radiacion ionizante a pacientes. Analisis en haz primario y secundario

Publishing Information

Imprint Title
Proceedings of the 5. Regional congress on radiation protection and safety; 2. Iberian and Latin American Congress on Radiological Protection Societies; Regional IRPA Congress
Imprint Pagination
[1776 p.]
Journal Page Range
[5 p.]

Conference

Title
5. Regional congress on radiation protection and safety; 2. Iberian and Latin American Congress on Radiological Protection Societies; Regional IRPA Congress
Dates
29 Apr - 4 May 2001
Place
Recife, PE (Brazil)

Optional Information

Notes
8 refs., 1 fig., 2 tabs., 2 graphs Imprint:Published only in CD-ROM. This electronic proceedings requires the Adobe Acrobat Reader 4.05 (included on this CD) and PC (at least 486 CPU) running Windows 95/98NT