Matrix-based image reconstruction methods for tomography
Description
Matrix methods of image reconstruction have not been used, in general, because of the large size of practical matrices, ill condition upon inversion and the success of Fourier-based techniques. An exception is the work that has been done at the Lawrence Berkeley Laboratory for imaging with accelerated radioactive ions. An extension of that work into more general imaging problems shows that, with a correct formulation of the problem, positron tomography with ring geometries results in well behaved matrices which can be used for image reconstruction with no distortion of the point response in the field of view and flexibility in the design of the instrument. Maximum Likelihood Estimator methods of reconstruction, which use the system matrices tailored to specific instruments and do not need matrix inversion, are shown to result in good preliminary images. A parallel processing computer structure based on multiple inexpensive microprocessors is proposed as a system to implement the matrix-MLE methods. 14 references, 7 figures
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS, PC A02; 3 as DE85003468.
Files
Additional details
Publishing Information
- Imprint Pagination
- 12 p.
- Report number
- LBL--17781
Conference
- Title
- Nuclear science symposium.
- Dates
- 31 Oct - 2 Nov 1984.
- Place
- Orlando, FL (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 16065007
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BEAM OPTICS; DATA PROCESSING; IMAGE PROCESSING; MATRICES; POSITRON COMPUTED TOMOGRAPHY
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; TOMOGRAPHY
Optional Information
- Secondary number(s)
- CONF-841007--31.