Extremity dosimeter studies

Description

Finger rings employing a single TLD-700 chip have served as the routine extremity dosimeter for the past several years. The response characteristics of the finger ring are described in this study. Proper annealing is crucial to accurate and precise readout and dose assignation. The standard PNL annealing procedure of 4000C for one hour followed by 1000C for two hours appears best; reader annealing is unacceptable, showing extremely poor precision. Although the rings are in the main suitable from a human factors standpoint, adjustable size is suggested for future models. Also, it should be noted that the rings are not suitable for neutron dosimetry, but can be used for beta dosimetry with a suitable correction factor

Files

12619508.pdf

Files (371.9 kB)

Name Size Download all
md5:db123f739fad0b3f7c9e5825990eb87a
371.9 kB Preview Download

Additional details

Publishing Information

Imprint Title
Hanford Personnel Dosimeter supporting studies FY-1980
Journal Page Range
p. 9.1-9.11.
Report number
PNL--3536