Published November 2005 | Version v1
Miscellaneous Open

Comparisons of the standards for air kerma of the PTB and the BIPM for 60Co and 137Cs gamma radiation

  • 1. Bureau International des Poids et Mesures (BIPM), 92 - Sevres (France)
  • 2. Physikalisch-Technische Bundesanstalt, Braunschweig (Germany)

Description

Direct comparisons of the standards for air kerma of the Physikalisch-Technische Bundesanstalt (PTB, Germany) and of the Bureau International des Poids et Mesures (BIPM) were carried out in the 60Co and 137 Cs radiation beams of the BIPM in 2000. The results, expressed as ratios of the PTB and the BIPM standards for air kerma, indicate a relative difference in 60Co of 9.9 x 10-3 with a combined standard uncertainty of 1.8 x 10-3, and in 137Cs of 6.4 x 10-3 with a combined standard uncertainty of 2.8 x 10-3. The earlier comparisons in 60Co γ rays made in 1971 (direct) and 1989 (indirect) resulted in an agreement of the two standards within 2 x 10-3. The differences obtained now are due to the application of new correction factors for wall effects and point source non-uniformity of the beam, kwall and kpn, for the PTB standards, which were calculated using Monte Carlo methods. (authors)

Availability note (English)

Available from INIS in electronic form

Files

38101988.pdf

Files (325.5 kB)

Name Size Download all
md5:1cc7f4b96f9bbd429880da338b15fe15
325.5 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
17 p.
Report number
BIPM--05-10

Optional Information

Notes
30 refs.