Performance analysis of a dedicated breast MR-HIFU system for tumor ablation in breast cancer patients
Creators
- 1. Imaging Division, University Medical Center Utrecht, Utrecht (Netherlands)
- 2. Philips Healthcare, Vantaa (Finland)
Description
MR-guided HIFU ablation is a promising technique for the non-invasive treatment of breast cancer. A phase I study was performed to assess the safety and treatment accuracy and precision of MR-HIFU ablation in breast cancer patients () using a newly developed MR-HIFU platform dedicated to applications in the breast. In this paper a technical analysis of the performance of the dedicated breast MR-HIFU system during breast tumors ablation is described. The main points of investigation were the spatial targeting accuracy and precision of the system and the performance of real-time respiration-corrected MR thermometry.
The mean targeting accuracy was in the range of 2.4–2.6 mm, whereas the mean targeting precision was in the range of 1.5–1.8 mm. To correct for respiration-induced magnetic field fluctuations during MR temperature mapping a look-up-table (LUT)-based correction method was used. An optimized procedural sedation protocol in combination with the LUT-based correction method allowed for precise MR thermometry during the ablation procedure (temperature standard deviation <3 °C). No unwanted heating in the near field (i.e. skin) nor in the far field (pectoral muscle) was detected.
The newly developed dedicated breast MR-HIFU system allows for safe, accurate and precise ablation of breast tumors. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/60/14/5527Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 60
- Journal Issue
- 14
- Journal Page Range
- p. 5527-5542
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 51040567
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ABLATION; ACCURACY; CORRECTIONS; HEATING; MAGNETIC FIELDS; MAMMARY GLANDS; MUSCLES; NEOPLASMS; PATIENTS; PERFORMANCE; RESPIRATION; SAFETY STANDARDS; SKIN
- Descriptors DEC
- BODY; DISEASES; GLANDS; ORGANS; STANDARDS