Published January 2000
| Version v1
Journal article
A pilot study on the quality control of film processing in medical radiology laboratories in Greece
Description
The results of a pilot study on the quality of film processing in 80 medical diagnostic radiology laboratories all over Greece are presented. The sensitometric technique for the evaluation of processing has been used to calculate film's base+fog, maximum optical density, speed and contrast, parameters which describe the performance characteristics of automatic film processors and films. The mean values of the base+fog and the maximum optical density were well within the acceptance limits. The film speed was almost constant while the film contrast showed significant variation
Additional details
Identifiers
- PII
- S0720048X99000765;
Publishing Information
- Journal Title
- European Journal of Radiology
- Journal Volume
- 33
- Journal Issue
- 1
- Journal Page Range
- p. 24-31
- ISSN
- 0720-048X
- CODEN
- EJRADR
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34033411
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; GREECE; IMAGE PROCESSING; PHOTOGRAPHIC FILMS; QUALITY CONTROL
- Descriptors DEC
- CONTROL; DEVELOPING COUNTRIES; DIAGNOSTIC TECHNIQUES; EUROPE; MEDICINE; NUCLEAR MEDICINE; PROCESSING; RADIOLOGY; WESTERN EUROPE
Optional Information
- Copyright
- Copyright (c) 2000 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.