Published January 2000 | Version v1
Journal article

A pilot study on the quality control of film processing in medical radiology laboratories in Greece

Description

The results of a pilot study on the quality of film processing in 80 medical diagnostic radiology laboratories all over Greece are presented. The sensitometric technique for the evaluation of processing has been used to calculate film's base+fog, maximum optical density, speed and contrast, parameters which describe the performance characteristics of automatic film processors and films. The mean values of the base+fog and the maximum optical density were well within the acceptance limits. The film speed was almost constant while the film contrast showed significant variation

Additional details

Identifiers

PII
S0720048X99000765;

Publishing Information

Journal Title
European Journal of Radiology
Journal Volume
33
Journal Issue
1
Journal Page Range
p. 24-31
ISSN
0720-048X
CODEN
EJRADR

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34033411
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; GREECE; IMAGE PROCESSING; PHOTOGRAPHIC FILMS; QUALITY CONTROL
Descriptors DEC
CONTROL; DEVELOPING COUNTRIES; DIAGNOSTIC TECHNIQUES; EUROPE; MEDICINE; NUCLEAR MEDICINE; PROCESSING; RADIOLOGY; WESTERN EUROPE

Optional Information

Copyright