Published 1989 | Version v1
Book

Dose reduction in fluoroscopic and interventional radiologic procedures

Creators

Description

This paper reports on a search for ways to reduce radiation exposure in fluoroscopy and interventional radiologic procedures. The authors have evaluated various parameters, shielding devices, and exposure procedures. In particular, they have studied the effects of various filters. It is possible to design a special filter that can modify the low-energy part of the x-ray spectrum, reducing both the patient's exposure and scattered radiation by 30%-40% without losing image quality. Using special phantoms and the experimental filter, they have determined quantitatively the changes in the contrast-detail curve, as well as the slight increase in noise

Additional details

Publishing Information

Publisher
Radiological Society of North America Inc.
Imprint Place
Oak Brook, IL (USA)
Imprint Title
Proceedings of the 75th anniversary scientific assembly and annual meeting Radiological Society of North America (Abstracts)
Imprint Pagination
654 p.
Journal Page Range
p. 428.

Conference

Title
75. anniversary scientific assembly and annual meeting of the Radiological Society of North America.
Dates
26 Nov - 1 Dec 1989.
Place
Chicago, IL (USA).

Optional Information

Secondary number(s)
CONF-8911163--.