Published November 2016 | Version v1
Journal article

Responsivity calibration of the LoWEUS spectrometer

  • 1. Space Sciences Laboratory, University of California, Berkeley, California 94720 (United States)
  • 2. Physics Division, Lawrence Livermore National Laboratory, Livermore, California 94550 (United States)
  • 3. Princeton Plasma Physics Laboratory, Princeton, New Jersey 08543 (United States)

Description

We performed an in situ calibration of the relative responsivity function of the Long-Wavelength Extreme Ultraviolet Spectrometer (LoWEUS), while operating on the Lithium Tokamak Experiment (LTX) at Princeton Plasma Physics Laboratory. The calibration was accomplished by measuring oxygen lines, which are typically present in LTX plasmas. The measured spectral line intensities of each oxygen charge state were then compared to the calculated emission strengths given in the CHIANTI atomic database. Normalizing the strongest line in each charge state to the CHIANTI predictions, we obtained the differences between the measured and predicted values for the relative strengths of the other lines of a given charge state. We find that a 3rd degree polynomial function provides a good fit to the data points. Our measurements show that the responsivity between about 120 and 300 Å varies by factor of ∼30.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
87
Journal Issue
11
Journal Page Range
vp.
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2016 Author(s)