Published June 2017
| Version v1
Journal article
GM tube radiation detector based on Time-to-Count method
- 1. China Institute for Radiation Protection, Taiyuan (China)
Description
The design of a new type of GM Tube radiation detector was introduced. The detector was designed according to the principle of Time-to-Count method. The hardware structure and program flow design of the radiation detector were presented. The test results showed that the linearity of Time-to-Count method is better than pulse counting method. Besides, the range upper limit based on Time-to-Count method extended two orders of magnitudes compared with the traditional pulse counting method. (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 37
- Journal Issue
- 6
- Journal Page Range
- p. 641-646
- ISSN
- 0258-0934
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 53032779
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; DESIGN; GEIGER-MUELLER COUNTERS; PULSES; RADIATION DETECTORS; TIME DEPENDENCE
- Descriptors DEC
- EVALUATION; MEASURING INSTRUMENTS; RADIATION DETECTORS
Optional Information
- Notes
- 4 figs., 6 tabs., 4 refs.