Published 2021 | Version v1
Journal article

Comparative evaluation of the performance of thin diodes used as on-line dosimeters in radiation processing applications

  • 1. Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN/SP), São Paulo, SP (Brazil)
  • 2. Universidade de São Paulo (IF/USP), São Paulo, SP (Brazil)

Description

In this work, we report a comparison between the performance of two samples of commercial PIN photodiodes (SFH206K from Osram® and S2506-04 from Hamamatsu®) mainly addressing the variation of their current sensitivities with accumulated dose ranging from 0-15 kGy. All the results so far obtained have revealed that the radiation induced currents are linearly dependent on dose rates from 3.65 to 55.64 Gy/h. The current sensitivity of both unirradiated diodes (0.178 nA.h/Gy.mm3) slightly decreases with accumulated dose, namely 0.32%/kGy (SFH206K) and 1.4%/kGy (S2506-04). Although the SFH206K device compares favorably with the S2506-04, both diodes can be considered as a low budge alternative, good enough for on-line dosimetry applications in the field of radiation processing. (author)

Additional details

Publishing Information

Journal Title
Brazilian Journal of Radiation Sciences
Journal Volume
9
Journal Issue
1A
Journal Page Range
8 p.
ISSN
2319-0612

Conference

Title
14. meeting on nuclear applications - ENAN
Dates
21-25 Oct 2019
Place
Santos, SP (Brazil)