Published May 1, 1994 | Version v1
Journal article

Determination of detection efficiency for radon and radon daughters with CR 39 track detector - a Monte Carlo study

Creators

  • 1. Faculty of Natural Sciences, Kragujevac (Yugoslavia)

Description

The detection effciency, ρ, (or a calibration coefficient k) for radon measurements with a solid state nuclear track detector CR 39 was determined by many authors. There is a considerable discrepancy among reported values for ρ. This situation was a challenge to develop a software program to calculation ρ. This software is based on Bethe-Bloch's expression for the stopping power for heavy charged particles in a medium, as wll as on the Monte Carlo Method. Track parameters were calculated by using an iterative procedure as given in G. Somogyi et al., Nucl. Instr. and Meth. 109 (1973) 211. Results for an open detector and for the detector in a diffusion chamber were presented in this article. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
344
Journal Issue
2
Journal Page Range
p. 406-414.
ISSN
0168-9002
CODEN
NIMAER