Transmission imaging with a coded source
- 1. Raytheon Research Divison, Waltham, MA
Description
The conventional approach to transmission imaging is to use a rotating anode x-ray tube, which provides the small, brilliant x-ray source needed to cast sharp images of acceptable intensity. Stationary anode sources, although inherently less brilliant, are more compatible with the use of large area anodes, and so they can be made more powerful than rotating anode sources. Spatial modulation of the source distribution provides a way to introduce detailed structure in the transmission images cast by large area sources, and this permits the recovery of high resolution images, in spite of the source diameter. The spatial modulation is deliberately chosen to optimize recovery of image structure; the modulation pattern is therefore called a ''code.'' A variety of codes may be used; the essential mathematical property is that the code possess a sharply peaked autocorrelation function, because this property permits the decoding of the raw image cast by th coded source. Random point arrays, non-redundant point arrays, and the Fresnel zone pattern are examples of suitable codes. This paper is restricted to the case of the Fresnel zone pattern code, which has the unique additional property of generating raw images analogous to Fresnel holograms. Because the spatial frequency of these raw images are extremely coarse compared with actual holograms, a photoreduction step onto a holographic plate is necessary before the decoded image may be displayed with the aid of coherent illumination
Files
Additional details
Publishing Information
- Imprint Title
- Proceedings of ERDA symposium on x- and gamma-ray sources and applications
- Imprint Pagination
- p. 133-136.
- Report number
- CONF-760539--
Conference
- Title
- ERDA symposium on x- and gamma-ray sources and applications.
- Dates
- 19 May 1976.
- Place
- Ann Arbor, Michigan, United States of America (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 8311232
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- HOLOGRAPHY; IMAGES; MODULATION; PHOTON TRANSMISSION SCANNING; X-RAY SOURCES
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; RADIATION SOURCES; X-RAY EQUIPMENT