Published December 2010 | Version v1
Journal article

Preparation of 32P radioactive covered metal stent: an experimental study in vitro

  • 1. Department of Radiology, the Affiliated Changzheng Hospital, No.2 Military Medical University, Shanghai (China)

Description

Objective: To optimize the preparation technique of 32P radioactive covered metal stent. Methods: The 32P radioactive material leakage rate in saline of polyvinyl alcohol membrane and polyurethane membrane was calculated respectively. The 32P radioactive material disposition homogeneity in different polyurethane membranes and in different portions of polyurethane membrane was determined separately. Polyurethane membrane homogeneity by drop-coating and dip-coating on metal stent was evaluated respectively. The results were statistically analyzed. Results: The 32P radioactive material leakage rate in saline of polyvinyl alcohol membrane was 90%, but the 32P radioactive material leakage rate of polyurethane membrane was very low. No significant difference in the 32P radioactive disposition homogeneity existed between different polyurethane membranes (Mann-Whitney Test, U = 10.0, P = 0.602 > 0.05), and the mean difference in different portion of polyurethane membrane was 11.76%. The polyurethane membrane on stent covered by dip-coating technique was homogeneous, but the membrane was not homogeneous,even with perforation,if covered by drop-coating technique. Conclusion: The polyurethane membrane on metallic stent which is covered by dip-coating technique can decrease the 32P radioactive material leakage in saline, and the distribution of the radioactive material in such a membrane is quite homogeneous. (authors)

Additional details

Publishing Information

Journal Title
Journal of Interventional Radiology
Journal Volume
19
Journal Issue
12
Journal Page Range
p. 968-971
ISSN
1008-794X

Optional Information

Notes
1 figs., 1 tabs., 16 refs.