The Joint Research Centre, in cooperation with the Commissariat a l'Energie Atomique et aux Energies Alternatives, produced a novel 243Am spike reference material for mass spectrometry. Americium solution with an isotopic composition of 88% 243Am and 12% 241Am was used as the source for the preparation of the spike material. The certified value of 5.696 (11) nmol g-1 for the amount content of 243Am and 0.136138 (54) for the n(241Am)/n(243Am) amount ratio were assigned. The assigned values from mass spectrometry were confirmed by alpha-particle spectrometry, alpha-particle counting at a defined solid angle, and high-resolution gamma-ray spectrometry. Furthermore, an external validation of the certified values was obtained from the results of an interlaboratory comparison exercise, using this americium reference solution as the test sample. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
327
Journal Issue
1
Journal Page Range
p. 495-504
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
43 refs.