Description
Charge-exchange spectroscopy on JET has become particularly challenging with the introduction of the ITER-like wall. The line intensities are weaker and contaminated by many nuisance lines. We have therefore upgraded the instrumentation to improve throughput and allow the simultaneous measurement of impurity and fuel-ion charge exchange by splitting the light between two pairs of imaging spectrometers using dichroic beam splitters. Imaging instruments allow us to stack 11 x 1 mm diameter fibres on the entrance slits without cross talk. CCD cameras were chosen to have 512 x 512 pixels to allow frame transfer times < 0.2 ms which with minimum exposure times of 5 ms give tolerable smearing even without a chopper. The image plane is optically demagnified 2:1 to match the sensor size of these cameras. Because the image plane of the spectrometer is tilted, the CCD must also be tilted to maintain focus over the spectrum (Scheimpflug condition). To avoid transverse keystoning (causing the vertical height of the spectra to change across the sensor), the configuration is furthermore designed to be telecentric by a suitable choice of the lens separation. The lens configuration is built almost entirely from commercial off-the-shelf components, which allowed it to be assembled and aligned relatively rapidly to meet the deadline for in-vessel calibration in the JET shutdown. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1063/1.5037639Additional details
Identifiers
- DOI
- 10.1063/1.5037639;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 89
- Journal Issue
- no.10
- Journal Page Range
- p. 1-5
- ISSN
- 0034-6748
INIS
- Country of Publication
- United States
- Country of Input or Organization
- France
- INIS RN
- 54072934
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- BEAMS; CALIBRATION; CHARGE EXCHANGE; CHARGE-COUPLED DEVICES; DESIGN; FIBERS; IMAGES; IONS; ITER TOKAMAK; JETS; LENSES; SENSORS; SPECTRA; SPECTROMETERS; SPECTROSCOPY
- Descriptors DEC
- CHARGED PARTICLES; CLOSED PLASMA DEVICES; MEASURING INSTRUMENTS; SEMICONDUCTOR DEVICES; THERMONUCLEAR DEVICES; THERMONUCLEAR REACTORS; TOKAMAK DEVICES; TOKAMAK TYPE REACTORS