Published October 1994 | Version v1
Journal article

Dosimetry in the moving gap region in craniospinal irradiation

  • 1. Arizona Univ., Tucson, AZ (United States)
  • 2. Fox Chase Cancer Center, Philadelphia, PA (United States). Dept. of Radiation Oncology

Description

To minimize the risk of overlap of adjoining orthogonal fields in craniospinal irradiation, a skin gap is usually maintained between the caudal margin of the brain fields and the cephalic margin of the spine field. The moving gap technique (feathering) is commonly used to improve the dose in the gap region. Using Kodak XV ready-pack films and a polystyrene phantom, the dosimetry in the moving gap region at the depth of spinal cord for different gap sizes and feathering step sizes was studied. The dose profiles in the moving gap region with half-beam block set-up were also measured. Our results show that the combination of a small skin gap and a larger feathering step size results in a higher dose in the moving gap compared with that identical gap and feathering step sizes. (author)

Additional details

Publishing Information

Journal Title
British Journal of Radiology
Journal Volume
67
Journal Issue
802
Journal Page Range
p. 1017-1022.
ISSN
0007-1285
CODEN