Published November 1982
| Version v1
Journal article
Scanning delay generator for measurement of kinetic decays using laser-induced fluorescence techniques
Creators
- 1. Department of Chemistry, Howard University, Washington, DC 20059
Description
An electronic device is described which generates a delayed pulse, where the delay is increased in a step-like manner. This device finds application in the measurement of kinetic decays, particularly when laser-induced fluorescent (LIF) detection of the decaying species is used. The circuit uses 15 low-power Schottky TTL chips and is contained within a dual width NIM module
Additional details
Publishing Information
- Journal Title
- Rev. Sci. Instrum.
- Journal Volume
- 53
- Journal Issue
- 11
- Series
- Rev. Sci. Instrum.
- Journal Page Range
- 1786-1788
- ISSN
- 0034-6748
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 14738633
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DELAY CIRCUITS; ELECTRONIC CIRCUITS; FLUORESCENCE; INTEGRATED CIRCUITS; LASER RADIATION; LOGIC CIRCUITS; PULSES; RADIATION EFFECTS; SPECIFICATIONS
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; LUMINESCENCE; RADIATIONS