Time-resolved dosimetry of pulsed electron beams in very high dose-rate, FLASH irradiation for radiotherapy preclinical studies

  • 1. Institut Curie, University Paris Saclay, PSL Research University, Inserm U 1021-CNRS UMR 3347, Centre Universitaire, 91405 Orsay Cedex (France)
  • 2. Institut Curie, Inserm U 1196 - CNRS UMR 9187, Bldg. 110, Centre Universitaire, 91405 Orsay Cedex (France)
  • 3. Institut Curie, Department of Translational Research, PSL Research University, Bldg. 110, Centre Universitaire, 91405 Orsay Cedex (France)
  • 4. Institut Curie, PSL Research University, Radiation Oncology Department, 26 rue d'Ulm, 75248 Paris Cedex 05 (France)

Description

Most anticancer radiation therapy facilities are based on linear electron accelerators with electron–photon conversion providing dose-rates in the range 0.03-0.40 Gy.s−1, and treatment plans usually involve daily fractions of 2 Gy cumulated for up to reaching a total dose close to the limit of tolerance of the normal tissues that surround tumors. We recently developed another methodology named "FLASH" that relies on very high dose-rate facilities and consists in delivering 10 Gy in a single microsecond pulse of relativistic electrons, or else in a limited number of pulses of 1-2 Gy each given in 100 ms temporal sequence. In mice FLASH was found to elicit a dramatic decrease of damage to normal tissues whilst keeping the anti-tumor efficiency unchanged. In the following we describe the methods used for beam monitoring in the FLASH mode with emphasis on techniques that provide proportional, time-resolved dosimetry of radiation at the submicrosecond time scale. These methods include measurement of the electron fluence, optically monitored chemical dosimeters in water, solid scintillation and Cerenkov light emission. An application to the calibration of Gafchromic™ films is described and the minimal requirements for dose monitoring in preclinical assays are discussed. Good repeatability and linearity of these techniques in a range of peak dose-rates from 2x102 to 4x107 Gy.s−1 and from 1 mGy to over 30 Gy per microsecond pulse have been obtained with an overall precision better than ± 2%.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2019.162537

Additional details

Identifiers

DOI
10.1016/j.nima.2019.162537;
PII
S0168900219310666;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
944
Journal Page Range
vp.
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
56005542
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
BEAM MONITORING; DOSE RATES; DOSES; DOSIMETRY; ELECTRON BEAMS; ELECTRONS; RADIOTHERAPY; SCINTILLATIONS; TIME RESOLUTION
Descriptors DEC
BEAMS; ELEMENTARY PARTICLES; FERMIONS; LEPTON BEAMS; LEPTONS; MEDICINE; MONITORING; NUCLEAR MEDICINE; PARTICLE BEAMS; RADIOLOGY; RESOLUTION; THERAPY; TIMING PROPERTIES

Optional Information

Copyright
Copyright (c) 2019 Elsevier B.V. All rights reserved.