Characterization of GE discovery IGS 740 angiography system by means of channelized Hotelling observer (CHO)

  • 1. Servizio di Fisica Medica, Azienda USL-IRCCS di Reggio Emilia, Reggio Emilia (Italy)
  • 2. Servizio di Gastroenterologia e Endoscopia Digestiva, Azienda USL-IRCCS di Reggio Emilia, Reggio Emilia (Italy)
  • 3. Servizio di Fisica Sanitaria, Azienda Ospedaliera Universitaria di Parma, Parma (Italy)

Description

The aim of the study is to use the well-known channelized Hotelling observer model (CHO) to characterize a recently installed angiography system (GE Discovery IGS 740) using sets of images of a contrast-detail phantom acquired with clinical protocols. A Leeds TO10 phantom was used. The phantom has 108 details: 12 diameters (size range: 0.25 mm–11 mm), each with nine contrasts (declared range: 0.012–0.930 at 70 kVp 1.00 with 1 mm Cu filtration). TO10 has been imaged between two 10 cm thick homogeneous solid water slabs. Two FOVs (32 cm and 20 cm) were used. Fluoroscopy images were taken using an abdominal protocol at two different frame rates (15 fps and 7.5 fps) and at two dose levels (low and normal); cineangiography images were acquired using an abdominal protocol at 15 fps at two dose levels (low and normal). A 40 Gabor channels CHO with internal noise was used. Human observers' studies were carried out to tune the internal noise parameter and to validate the model observer. Contrast-detail curves were obtained from the CHO output using a visibility threshold of 75% and fitted with Rose's model theory in order to characterize the angiography system. Wilcoxon rank-sum tests were performed to investigate possible differences among the different sets of images.

The CHO can distinguish between the two dose levels (p -values  <  0.002), while FOV and frame rate do not affect the contrast-detail curves significantly. It is important to note that the CHO does not find statistically significant differences between a fluoroscopy with FOV  =  20 cm at normal dose level (17.6 mGy min−1) and a cineangiography with FOV  =  32 cm low dose level (42.1 mGy min−1). This result can lead to a dose reduction of about 70% for our specific task (i.e. a static, disc shaped object at known location in homogeneous field). Given their stability in comparison to human observers, model observers provide an effective tool for image quality evaluation. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/ab144c

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
64
Journal Issue
9
Journal Page Range
[15 p.]
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52003803
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BLOOD VESSELS; COPPER; GERMANIUM; PHANTOMS; RADIATION DOSES
Descriptors DEC
BODY; CARDIOVASCULAR SYSTEM; DOSES; ELEMENTS; METALS; MOCKUP; ORGANS; STRUCTURAL MODELS; TRANSITION ELEMENTS