Published March 1978 | Version v1
Journal article

Csub(lambda) and Csub(E) values

  • 1. Cookridge Hospital, Leeds (UK)

Description

Csub(lambda) and Csub(E) values for use with a Farmer ionisation chamber were obtained with an MEL(SL 75/20) linear accelerator using both an absorbed dose calorimeter and ferrous sulphate. Results are summarised and compared with ICRU recommended values. They support the view that the discrepancy in the values of Csub(lambda) and Csub(E) is due to an error in the definition of Csub(lambda) rather than Csub(E). Equations are proposed for the calculation of C at all photon energies taking into consideration the nature and thickness of the chamber wall material

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine & Biology
Journal Volume
23
Journal Issue
2
Series
Phys. Med. Biol.
Journal Page Range
336-339
ISSN
0031-9155

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent