EPR imaging from projections: errors due to misalignment of projection centres and their rectification by a novel acquisition modality
Creators
- 1. INFM, c/o Dipartimento di Scienze e Tecnologie Biomediche, Universita degli Studi di L'Aquila, Via Vetoio 10, 67010 Coppito, L'Aquila (Italy).
- 2. INFM, c/o Centro Interdipartimentale di Risonanza Magnetica, Universita degli Studi di L'Aquila, Via Vetoio 10, 67010 Coppito, L'Aquila (Italy)
- 3. INFM, c/o Dipartimento di Scienze e Tecnologie Biomediche, Universita degli Studi di L'Aquila, Via Vetoio 10, 67010 Coppito, L'Aquila (Italy)
- 4. INFM, c/o Dipartimento di Ingegneria Elettrica, Universita degli Studi di L'Aquila, Poggio di Roio, 67100, Roio, L'Aquila (Italy)
Description
Continuous wave and pulsed wave electron paramagnetic resonance imaging (EPRI) makes use of classical methods of acquisition of projections. Acquisition/reconstruction techniques, such as spin-echo, gradient-echo, etc, cannot be applied to EPRI because they would require very short switching times for the gradient coils. Due to the use of the polar acquisition technique, it is necessary to define a centre of rotation about which the measured projections are rotated during the reconstruction process. This centre represents the point at which the field gradient coils must produce zero magnetic field. Due to the presence of a magnetic field control system that serves to compensate for field variations, principally due to heating, some interference can occur in the control system between the main magnetic field and the magnetic field produced by the gradient coils. The effect changes as the orientation changes. This results in a shift of the centres of the projections as a function of the variation of magnetic field produced by the gradient coils on the control Hall probe. If this condition is present, some artefacts can appear on the reconstructed image. This effect is irrelevant when EPR is used for imaging of paramagnetic probes whose linewidths are of the order of 10-4 T, while it can be significant in the case of linewidths of the order of 10-5 T or lower or when EPR is used in microimaging applications (i.e. for high values of magnetic field gradient). We describe the effects that misalignments of the projections have on the reconstructed images. We present a useful method for estimating the real position of the centre and correcting the measured projections before the application of the reconstruction algorithm. Moreover, we demonstrate the functioning of our technique by presenting some examples of EPR reconstruction collected by an X-band EPR imaging apparatus. (author)
Availability note (English)
Available online at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://www.iop.org/;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 45
- Journal Issue
- 11
- Journal Page Range
- p. 3135-3142
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43119805
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DIAGNOSTIC USES; ELECTRON SPIN RESONANCE; EPR SPECTROMETERS; IMAGE PROCESSING; SPIN ECHO
- Descriptors DEC
- MAGNETIC RESONANCE; MEASURING INSTRUMENTS; PROCESSING; RESONANCE; SPECTROMETERS; USES
Optional Information
- Notes
- Refs; This record replaces 31063644