Published July 2019
| Version v1
Journal article
Depth dose distribution measurements by X-ray CT. Comparison with and without tin filter
- 1. Niigata University Medical and Dental Hospital, Department of Radiology, Niigata (Japan)
- 2. Niigata University of Health and Welfare, Faculty of Medical Technology, Niigata (Japan)
- 3. Tohoku University, Graduate School of Medicine, Sendai, Miyagi (Japan)
Description
In recent years, X-ray CT apparatus has been applied clinically as Spectrum Shaping technology using tin filter has appeared as a new exposure reduction technique. In this study, the depth dose distribution inside the phantom was measured using a CTDI measurement phantom. Measurement using a tin filter showed lower measured values at any depth compared with other conditions. As a result, X-ray examination using a tin filter was considered to be effective for reducing patient exposure dose. (author)
Availability note (English)
Available from https://doi.org/10.18888/rp.0000000956Additional details
Additional titles
- Original title (Japanese)
- X線CT撮影による深部線量分布測定.錫フィルタの有無による比較
Identifiers
Publishing Information
- Journal Title
- Rinsho Hoshasen
- Journal Volume
- 64
- Journal Issue
- 8
- Series
- 雑誌名:臨床放射線
- Journal Page Range
- p. 1085-1092
- ISSN
- 0009-9252
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 51015042
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; DEPTH DOSE DISTRIBUTIONS; ELECTRIC POTENTIAL; FILTERS; INTERNAL IRRADIATION; MEDICAL EXAMINATIONS; MILLI GY RANGE 01-10; PHANTOMS; RADIATION DOSES; TIN; X RADIATION
- Descriptors DEC
- ABSORBED DOSE RANGE; DIAGNOSTIC TECHNIQUES; DOSES; ELECTROMAGNETIC RADIATION; ELEMENTS; IONIZING RADIATIONS; IRRADIATION; MEDICAL SURVEILLANCE; METALS; MILLI GY RANGE; MOCKUP; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSE RANGES; RADIATIONS; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes