Description

In recent years, X-ray CT apparatus has been applied clinically as Spectrum Shaping technology using tin filter has appeared as a new exposure reduction technique. In this study, the depth dose distribution inside the phantom was measured using a CTDI measurement phantom. Measurement using a tin filter showed lower measured values at any depth compared with other conditions. As a result, X-ray examination using a tin filter was considered to be effective for reducing patient exposure dose. (author)

Availability note (English)

Available from https://doi.org/10.18888/rp.0000000956

Additional details

Additional titles

Original title (Japanese)
X線CT撮影による深部線量分布測定.錫フィルタの有無による比較

Identifiers

Publishing Information

Journal Title
Rinsho Hoshasen
Journal Volume
64
Journal Issue
8
Series
雑誌名:臨床放射線
Journal Page Range
p. 1085-1092
ISSN
0009-9252