Comments on the measurement of line spread functions; and reply
Creators
- 1. Eastman Kodak Co., Rochester, NY (USA)
- 2. Tokoha Gakuen Univ., Sena, Shizuoka (Japan). Faculty of Education
- 3. Miyazaki Medical Coll. Hospital (Japan). Dept. of Radiology
- 4. Gifu Coll. of Medical Technology (Japan)
- 5. Gifu Univ. (Japan). Faculty of Engineering
Description
The authors take issue with several points in an article by Uchida et al (1986), suggesting an alternative method for measuring the line spread function of radiographic screen-film systems that avoids the necessity of combining partial LSF obtained from several levels of x-ray exposure. Uchida's reply is included. (U.K.)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 33
- Journal Issue
- 4
- Series
- Phys. Med. Biol.
- Journal Page Range
- 491-493
- ISSN
- 0031-9155
- CODEN
- PHMBA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 19067812
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DISPLAY DEVICES; ERRORS; IMAGE PROCESSING; IMAGES; PHOTOGRAPHIC FILMS; SPATIAL RESOLUTION; X RADIATION
- Descriptors DEC
- COMPUTER-GRAPHICS DEVICES; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; RADIATIONS; RESOLUTION