Full text: In the Tono Geoscience Center, Japan Atomic Energy Agency, investigation of deep underground environments for R&D program related to the geological disposal of High-Level Radioactive Waste (HLW) has been performed by using various dating systems including an AMS system. In response to the increasing demand for our AMS measurements especially from a newly established R&D program supporting development of technology for geological disposal of HLW, a state-of-the-art multi-nuclide AMS system was installed. This system is equipped with a 300 kV AMS (4103Bo-AMS), namely "JAEA-AMS-TONO- 300kV", manufactured by High Voltage Engineering, Europa B.V., Netherlands. The system has capability to measure four nuclides: carbon-14, beryllium-10, aluminium-26 and iodine-129. JAEAAMS-TONO-300kV is featuring the SO-110C ion source, a low energy spectrometer with electrostatic analyzer (ESA) and injection magnet with fast sequential isotope injection, tandem accelerator which maximum voltage is 300 kV, a high energy analyzer comprised of one ESA and two analysis magnets which degrees are 90 and 120, and an ionization chamber with a gas cell. The system footprint is 6.9 m x 5.0 m. The system structure and features, as well as the results of performance test will be presented. This installation was carried out by a contract with the Ministry of Economy, Trade and Industry of Japan as part of its R&D supporting program titled "Establishment of Advanced Technology for Evaluating the Long-term Geosphere Stability on Geological Disposal Project of Radioactive Waste (Fiscal Year 2020)".
Additional details
Publishing Information
- Imprint Title
- 15th International Conference on Accelerator Mass Spectrometry. Program and abstracts
- Imprint Pagination
- 303 p.
- Journal Page Range
- p. 300
- Report number
- INIS-AU--0116
Conference
- Title
- International Accelerator Mass Spectrometry Conference
- Acronym
- 2021 AMS-15
- Dates
- 15-19 Nov 2021
- Place
- Sydney, NSW (Australia)
INIS
- Country of Publication
- Australia
- Country of Input or Organization
- Australia
- INIS RN
- 54119916
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCELERATORS; ALUMINIUM 26; BERYLLIUM 10; CARBON 14; ELECTROSTATIC ANALYZERS; INSTALLATION; IODINE 129; ISOTOPES; JAEA; LABORATORY EQUIPMENT; MASS SPECTROSCOPY; MULTI-ELEMENT ANALYSIS
- Descriptors DEC
- ALKALINE EARTH ISOTOPES; ALUMINIUM ISOTOPES; BEAM ANALYZERS; BERYLLIUM ISOTOPES; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; CARBON ISOTOPES; CHEMICAL ANALYSIS; EQUIPMENT; EVEN-EVEN NUCLEI; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IODINE ISOTOPES; ISOTOPES; JAPANESE ORGANIZATIONS; LIGHT NUCLEI; NATIONAL ORGANIZATIONS; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIOISOTOPES; SECONDS LIVING RADIOISOTOPES; SPECTROSCOPY; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- Abstract only, full text in this record Imprint:Virtual conference. Includes obituary for Alan Williams (ANSTO Organising Committee) and 'In Memoriam' presentations for Professor Didier Bourl#Latin Small Letter E With Grave#s, Robert John 'Jack' Cornett and Ken Purser.