Is lead shielding of patients necessary during fluoroscopic procedures? A study based on kyphoplasty
- 1. University of Colorado School of Medicine, Department of Radiology, Aurora, CO (United States)
Description
To determine the benefits, risks, and limitations associated with wrapping a patient with lead shielding during fluoroscopy-guided kyphoplasty procedures as a way to reduce operator radiation exposure. An anthropomorphic phantom was used to mimic a patient undergoing a kyphoplasty procedure under fluoroscopic guidance. Radiation measurements of the air kerma rate (AKR) were made at several locations and under various experimental conditions. First, AKR was measured at various angles along the horizontal plane of the phantom and at varying distances from the phantom, both with and without a lead apron wrapped around the lower portion of the phantom (referred to here as phantom shielding). Second, the effect of an operator's apron was simulated by suspending a lead apron between the phantom and the measurement device. AKR was measured for the four shielding conditions - phantom shielding only, operator apron only, both phantom shielding and operator apron, and no shielding. Third, AKR measurements were made at various heights and with varying C-arm angle. At all locations, the phantom shielding provided no substantial protection beyond that provided by an operator's own lead apron. Phantom shielding did not reduce AKR at a height comparable to that of an operator's head. Previous reports of using patient shielding to reduce operator exposure fail to consider the role of an operator's own lead apron in radiation protection. For an operator wearing appropriate personal lead apparel, patient shielding provides no substantial reduction in operator dose. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00256-017-2756-9Additional details
Identifiers
Publishing Information
- Journal Title
- Skeletal Radiology
- Journal Volume
- 47
- Journal Issue
- 1
- Journal Page Range
- p. 37-43
- ISSN
- 0364-2348
- CODEN
- SKRADI
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 49021166
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BONE FRACTURES; COMPARATIVE EVALUATIONS; FLUOROSCOPY; KERMA; LEAD; MEASURING METHODS; MEDICAL PERSONNEL; OCCUPATIONAL EXPOSURE; OCCUPATIONAL SAFETY; PHANTOMS; PLASTIC SURGERY; PROTECTIVE CLOTHING; RADIATION DOSES; RADIATION PROTECTION; SHIELDING; SHIELDING MATERIALS; SIMULATION; VERTEBRAE
- Descriptors DEC
- BIOMEDICAL RADIOGRAPHY; BODY; CLOTHING; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; ELEMENTS; EVALUATION; INJURIES; MATERIALS; MEDICINE; METALS; MOCKUP; NUCLEAR MEDICINE; ORGANS; PERSONNEL; RADIOLOGY; SAFETY; SKELETON; STRUCTURAL MODELS; SURGERY