In order to carry out thoron sensitivity test for passive radon detectors, a thoron chamber system has been set up at NIRS, Japan. The thoron chamber system consists of four components: the exposure, monitoring, calibration, and humidity control systems, which was mounted in this study due to humidity dependence on the thoron concentration emanated from lantern mantles as the thoron source. The thoron concentration in the thoron chamber is controlled by humidity passed through the thoron source and the weight of the lantern mantle