Description

The problem of preparing materials of biological origin for trace element analysis using PIXE (proton-induced X-ray emission) is examined. A distinction is made between targets in which beam energy loss and X-ray absorption are negligible (thin) and those in which they are not (thick). The computer code FUDGE which calculates X-ray absorption factors for thick targets is described and the results of calculations for several organic materials are presented. The results of analyses of both thick and thin targets of NBS SRM 1577 (bovine liver) are reported, and their accuracy, precision and detection limits are compared. (Auth.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods
Journal Volume
142
Journal Issue
1-2
Series
Nucl. Instrum. Methods.
Journal Page Range
67-77

Conference

Title
International conference on particle induced X-ray emission and its analytical applications.
Dates
23 - 26 Aug 1976.
Place
Lund, Sweden.