Published December 1997
| Version v1
Report
The quality assurance regime in fabrication and calibration of dose rate meters with Geiger-Mueller counters
Description
EG and G Berthold use Geiger-Mueller counters bought elsewhere for fabrication of their hand-held or stationary probes for dose rate measurement. The quality assurance regime requires for some of the detectors an input performance test, the oscillograph showing the pulse shapes under various conditions. The tests to be performed with the Geiger-Mueller counters are roughly the same as those required for proportional counters. (orig./CB)
Abstract (German)
EG und G Berthold stellt selbst keine Geiger/Mueller-Zaehlrohre her, baut aber zugekaufte GM-Detektoren in Handgeraete oder stationaere Sonden zur DL-Messung ein. Bestimmte Typen dieser Detektoren unterliegen im Rahmen der Qualitaetssicherung einer Eingangspruefung, bei der am Oszillograph unter verschiedenen Bedingungen die Impulsformen kontrolliert werden. Bei der Montage und Kalibrierung der Geraete bzw. Sonden finden im wesentlichen die gleichen Kontrollen wie bei Proportionalzaehlrohrsystemen statt. Es sind dies Plateauaufnahme und Kontrolle des Arbeitspunkts, Kontrolle des Kalibrierfaktors, Bestimmung der Eigennulleffektsrate, und stichprobenweise Plateauaufnahmen und Dauerpruefungen bei extremen Temperaturen (+50 C und -30 C). (orig.)Additional details
Additional titles
- Original title (German)
- Qualitaetssicherung bei der Herstellung und Kalibrierung von Dosisleistungsmessern mit Geiger-Mueller-Zaehlrohren
Publishing Information
- ISBN
- 3-89701-054-2
- Imprint Title
- 110. PTB seminar: Dose rate measurements of ionizing radiation in the range of natural ambient radiation. Proceedings
- Imprint Pagination
- 192 p.
- Journal Page Range
- p. 42-45
- ISSN
- 0172-7095
- Report number
- PTB-Dos--26
Conference
- Title
- Dose rate measurements of ionizing radiation in the range of natural ambient radiation
- Original Conference Title
- 110. PTB-Seminar: Ortsdosismessungen Ionisierender Strahlung im Bereich der Natuerlichen Umgebungsstrahlung
- Acronym
- 110. PTB seminar
- Dates
- 30 Nov - 1 Dec 1993
- Place
- Braunschweig (Germany)
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 30005846
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- ACCURACY; CALIBRATION; DOSIMETRY; EFFICIENCY; GEIGER-MUELLER COUNTERS; MANUFACTURING; QUALITY ASSURANCE
- Descriptors DEC
- MEASURING INSTRUMENTS; RADIATION DETECTORS
Optional Information
- Notes
- Imprint:110. PTB-Seminar: Ortsdosisleistungsmessungen ionisierender Strahlung im Bereich der natuerlichen Umgebungsstrahlung. Vortraege