Published December 1997 | Version v1
Report

The quality assurance regime in fabrication and calibration of dose rate meters with Geiger-Mueller counters

Creators

  • 1. EG und G Berthold, Bad Wildbad (Germany)

Description

EG and G Berthold use Geiger-Mueller counters bought elsewhere for fabrication of their hand-held or stationary probes for dose rate measurement. The quality assurance regime requires for some of the detectors an input performance test, the oscillograph showing the pulse shapes under various conditions. The tests to be performed with the Geiger-Mueller counters are roughly the same as those required for proportional counters. (orig./CB)

Abstract (German)

EG und G Berthold stellt selbst keine Geiger/Mueller-Zaehlrohre her, baut aber zugekaufte GM-Detektoren in Handgeraete oder stationaere Sonden zur DL-Messung ein. Bestimmte Typen dieser Detektoren unterliegen im Rahmen der Qualitaetssicherung einer Eingangspruefung, bei der am Oszillograph unter verschiedenen Bedingungen die Impulsformen kontrolliert werden. Bei der Montage und Kalibrierung der Geraete bzw. Sonden finden im wesentlichen die gleichen Kontrollen wie bei Proportionalzaehlrohrsystemen statt. Es sind dies Plateauaufnahme und Kontrolle des Arbeitspunkts, Kontrolle des Kalibrierfaktors, Bestimmung der Eigennulleffektsrate, und stichprobenweise Plateauaufnahmen und Dauerpruefungen bei extremen Temperaturen (+50 C und -30 C). (orig.)
Part of:
110. PTB seminar: Dose rate measurements of ionizing radiation in the range of natural ambient radiation. Proceedings

Additional details

Additional titles

Original title (German)
Qualitaetssicherung bei der Herstellung und Kalibrierung von Dosisleistungsmessern mit Geiger-Mueller-Zaehlrohren

Publishing Information

ISBN
3-89701-054-2
Imprint Title
110. PTB seminar: Dose rate measurements of ionizing radiation in the range of natural ambient radiation. Proceedings
Imprint Pagination
192 p.
Journal Page Range
p. 42-45
ISSN
0172-7095
Report number
PTB-Dos--26

Conference

Title
Dose rate measurements of ionizing radiation in the range of natural ambient radiation
Original Conference Title
110. PTB-Seminar: Ortsdosismessungen Ionisierender Strahlung im Bereich der Natuerlichen Umgebungsstrahlung
Acronym
110. PTB seminar
Dates
30 Nov - 1 Dec 1993
Place
Braunschweig (Germany)

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
30005846
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; CALIBRATION; DOSIMETRY; EFFICIENCY; GEIGER-MUELLER COUNTERS; MANUFACTURING; QUALITY ASSURANCE
Descriptors DEC
MEASURING INSTRUMENTS; RADIATION DETECTORS

Optional Information

Notes
Imprint:110. PTB-Seminar: Ortsdosisleistungsmessungen ionisierender Strahlung im Bereich der natuerlichen Umgebungsstrahlung. Vortraege